Meet the requirements of general safety standards such as CCC, IEC, EN, VDE, BS, U'er, JIS, etc;
Adopting a 4.3-inch TFT-LCD display with simple panel operation;
Voltage ramp up and ramp down function. The waiting time for testing insulation resistance can be set arbitrarily;
Pre set upper and lower limits, intelligent judgment, sound and light alarm;
The DC withstand voltage test has a fast discharge function to avoid the risk of electric shock caused by residual voltage on the tested object;
Arc detection function;
Anti electric wall technology can detect circuits at any time, prevent electric shock injuries, and ensure personnel safety;
Overvoltage and overcurrent protection, current base reset function;
Can store 100 test files, each file can store 20 test steps, and test modes can be combined arbitrarily;
Rich interfaces, supporting both CPI and MODBUS protocols, making it easy to build automated testing systems;
Comes with 4-channel and 8-channel dual channel scanning interfaces;
Contact inspection function
Automated testing system
Lighting industry
household appliances
new energy vehicle
Transformers and motors
Electronic components
electrical equipment
medical equipment
heating appliance
photovoltaic industry
| model | HY9320-S4A | HY9320-S8A | |
| Scan interface | 4-way dual end | 8-way dual end | |
| Contact inspection | have | have | |
| Withstand Voltage Test | |||
| output voltage | AC | 0.050kV—5.000kV ±(1.0% Set +5V) | |
| DC | 0.050kV—6.000kV ±(1.0% Set +5V)) | ||
| Current testing range | AC | 0.001mA – 20.00mA ±(2.0% reading +5 characters) | |
| DC | 0.1uA – 10.00mA ±(2.0% reading +5 characters) | ||
| Quick discharge function | Automatic discharge after testing (DCW) | ||
| Insulation resistance test | |||
| output voltage | 0.050kV -2.500kV (optional 5.00kV) | ||
| Resistance testing range | 0.1MΩ– 100GΩ | ||
| Resistance testing accuracy | ≥500V 1MΩ– 1GΩ ±(5% reading +5 characters) 1G Ω– 10G Ω± (10% reading+5 words) 10G Ω– 50G Ω± (15% reading+5 words) 50G Ω -100G Ω for reference only, no accuracy requirements < 500V0.1M Ω– 1G Ω± (10% reading+5 words) 1G Ω -10G Ω for reference only, no accuracy requirements | ||
| Discharge function | Automatic discharge after the test is completed | ||
| ARC Detector | |||
| measurement range | AC | Levels 1-9 Corresponding currents: 20mA, 18mA, 16mA, 14mA, 12mA, 10mA, 7.7mA, 5.5mA, 2.8mA | |
| DC | Levels 1-9 | ||
| Parameter Settings | |||
| Voltage rise time | 0.1s – 999.9s | ||
| Test time setting | 0.1s -999.9s (when TIMER ON) | ||
| Voltage drop time | 0 s – 999.9s, (Only after passing the withstand voltage) | ||
| Voltage waiting time | 0.3s – 999.9s (Only DC withstand voltage, and it meets the requirement that the rise time + test time > waiting time) | ||
| time accuracy | ± (0.2% set value ± 0.1s) | ||
| memory | |||
| memory | Programmable 100 test files, each file can have 20 test items. | ||
| Ordering Information | |||
| Standard Accessories | HY900R: A pair of high-voltage test clips HY900B: High Voltage Test Rod HY106: RS232 communication cable HY917A: Voltage withstand test line X2 HY913A: Voltage withstand test line X4 (only HY9320-S4A) HY913A: Voltage withstand test line X8 (only HY9320-S8A) | ||